Cat. No. 04-270-0
Plasma coagulation measurement such as PT, PTT, FIB, and etc. as well as regular biochemical analysis is available by AE-600N!
| Filter | Interference filter (340, 405, 505, 546, 576, 630 nm. 1 other filter is available as option) |
|---|---|
| Filter accuracy | ±2% |
| Filter wavelength width | 10 nm (1/2 T%) |
| Measurement mode | Abs, End point, Kinetic, 2 point method, lasma coagulation |
| Optical measurement method | Single wavelength or dual wavelength |
| Display | LCD 320 x 240 matirix |
| Printer | Thermal method(58 mm width printer paper) |
| Abs measurement range | 0-2.5Abs |
| Sample holder | 30 μL |
| Sample setting volume | 200~999µL |
| Measurement parameter memory | 120 items |
| Stability | 0.003Abs/hour (10 min. after Power ON) |
| Temperature accuracy | ±0.1℃(25℃、30℃、32℃、37℃) |
| Interface | RS-232C |
| Power supply | 110/220V ±10% |
| Dimension | 440W x 345L x 190 H |
| Weight | 6 kg NET |
| Standard Accessories | Amount |
|---|---|
| Lamp | 1 |
| Fuse | 2 |
| Roller pump tube | 3 |
| Sample suction tube | 1 |
| Drain tube | 1 |
| Printer paper | 2 |
| Coagulation holder | 1 |
| Coagulation test tube | 100 |
| Cleaning solution | 1 |
| Dust cover | 1 |
| Operation manual | 1 |
Specifications subject to change without notice.
各製品の外観、仕様等は改良のため予告なく変更する場合があります。あらかじめご了承ください。