features

  1. Accurate measurement data by 6 interference filters of excellent quality
  2. Max. 96 parameters can be input in memory
  3. Standard graph in display enables easy correction
  4. Reaction process in kinetic measurement observed in display
  5. Temperature of sample holder controlled by Peltier element, enabling temperature setting at 25℃, 30℃, 32℃, and 37℃
  6. PT, PTT, FIB, also factor measurement available in plasma coagulation measurement
  7. Fine quality of micro flow cell and short-lengthened sample suction tube reduce influence of carry-over
  8. External devices such as PC are applicable by RS-232C

▲ Top of page

specification

Specification of AE-600N
Filter Interference filter
(340, 405, 505, 546, 576, 630 nm.
1 other filter is available as option)
Filter accuracy ±2%
Filter wavelength width 10 nm (1/2 T%)
Measurement mode Abs, End point, Kinetic, 2 point method, lasma coagulation
Optical measurement method Single wavelength or dual wavelength
Display LCD 320 x 240 matirix
Printer Thermal method(58 mm width printer paper)
Abs measurement range 0-2.5Abs
Sample holder 30 μL
Sample setting volume 200~999µL
Measurement parameter memory 120 items
Stability 0.003Abs/hour (10 min. after Power ON)
Temperature accuracy ±0.1℃(25℃、30℃、32℃、37℃)
Interface RS-232C
Power supply 110/220V ±10%
Dimension 440W x 345L x 190 H
Weight 6 kg NET

▲ Top of page

accessories

accessories of AE-600N
Standard AccessoriesAmount
Lamp 1
Fuse 2
Roller pump tube 3
Sample suction tube1
Drain tube 1
Printer paper 2
Coagulation holder 1
Coagulation test tube100
Cleaning solution 1
Dust cover 1
Operation manual 1

▲ Top of page


Specifications subject to change without notice.

各製品の外観、仕様等は改良のため予告なく変更する場合があります。あらかじめご了承ください。